Upgraded Cobas Ampliprep-Cobas TaqMan version 2.0 HIV-1 RNA quantification assay versus first version: correction of underestimations.

Wirden M, Tubiana R, Fourati S, Thevenin M, Simon A, Canestri A, Ait-Arkoub Z, Soulie C, Marcelin AG, Katlama C, Calvez V.

J Clin Microbiol. 2011 Jul;49(7):2700-2. doi: 10.1128/JCM.00061-11. Epub 2011 May 4.

The large underestimations of HIV RNA quantification observed in 17 patients with the first version of Cobas TaqMan assay have been successfully corrected in the upgraded version 2.0. In comparison with the Abbott RealTime assay, the mean difference that was 1.18 log(10) copies/ml is now zero. The discrepancies have disappeared.